天天摸天天碰天天添中文字幕,国产区点击进入在线影院,无码人妻一区二区免费AV,亚洲精品国产av现线

產(chǎn)品中心

人白介素6(IL6)酶聯(lián)免疫吸附檢測(cè)試劑盒
ELK1156
規(guī)格: 價(jià)格:
48T ¥1680.00
96T ¥2400.00

Overview 文獻(xiàn)

Product name: Human IL6(Interleukin 6) ELISA Kit
Reactivity: Human
Alternative Names: IL-6;MGI2-A; MGI2A; HGF; BSF2; HSF; IFNB2; B-Cell Stimulatory Factor-2; Hybridoma/Plasmacytoma Growth Factor; Hepatocyte Stimulating Factor; Cytotoxic T-Cell Differentiation Factor
Assay Type: Sandwich
Sensitivity: 3.3 pg/mL
Standard: 500 pg/mL
Detection Range: 7.82-500 pg/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Cytokine;Tumor immunity;Infection immunity;Cardiovascular biology;
Uniprot ID: P05231
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human IL6. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human IL6. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human IL6, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human IL6 in the samples is then determined by comparing the OD of the samples to the standard curve.

標(biāo)準(zhǔn)曲線(xiàn)

Concentration (pg/mL) OD Corrected OD
500.00 2.328 2.260
250.00 2.045 1.977
125.00 1.412 1.344
62.50 0.780 0.712
31.25 0.444 0.376
15.63 0.262 0.194
7.82 0.132 0.064
0.00 0.068 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant IL6 and the recovery rates were calculated by comparing the measured value to the expected amount of IL6 in samples.
Matrix Recovery range Average
serum(n=5) 97-105% 101%
EDTA plasma(n=5) 87-99% 92%
Heparin plasma(n=5) 88-107% 94%

線(xiàn)性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IL6 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 86-93% 91-103% 87-98% 92-99%
EDTA plasma(n=5) 89-99% 83-96% 98-104% 88-101%
Heparin plasma(n=5) 85-102% 84-99% 97-101% 89-102%