Product name: EasyStep Human HP(Haptoglobin) ELISA Kit
Reactivity: Human
Alternative Names: HP; Hp2-Alpha; Alpha-2-Macroglobulin; Zonulin; Hpt; Haptoglobin
Assay Type: Sandwich
Sensitivity: 0.02 ng/mL
Standard: 400 ng/mL
Detection Range: 6.25-400 ng/mL
Sample Type: Serum, plasma
Assay Length: 1.5h
Research Area: Infection immunity;Hematology;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human HP, and the Human HP standard plate wells that pre-coated using protein-related techniques are provided separately. Standard/Sample Diluent Buffer or samples are added to the appropriate microtiter plate wells ,then added a HRP-conjugated antibody specific to Human HP. After TMB substrate solution is added, only those wells that contain Human HP and HRP-conjugated antibody will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human HP in the samples is then determined by comparing the OD of the samples to the standard curve.

標(biāo)準(zhǔn)曲線

Concentration (ng/mL) OD Corrected OD
400.00 1.997 1.912
200.00 1.629 1.544
100.00 1.194 1.109
50.00 0.971 0.886
25.00 0.509 0.424
12.50 0.364 0.279
6.25 0.219 0.134
0.00 0.085 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant HP and the recovery rates were calculated by comparing the measured value to the expected amount of HP in samples.
Matrix Recovery range Average
serum(n=5) 80-95% 87%
EDTA plasma(n=5) 80-93% 86%
Heparin plasma(n=5) 95-107% 101%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of HP and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 86-93% 91-105% 87-99% 99-105%
EDTA plasma(n=5) 88-95% 96-102% 89-103% 89-96%
Heparin plasma(n=5) 87-101% 87-96% 82-93% 89-97%